UV and fluorescence detectors are widely used. Therefore ion-pair reagents must lack UV absorption and fluorescence themselves to obtain highly sensitive detection of samples. The UV absorption of sodium alkanesulfonates and quaternary ammonium salts is minimal so that these reagents can be used for reliable HPLC analysis. On the other hand, when a sample lacks sufficient UV absorption or fluorescence, the use of IPA-DAS [A5701] allows for high-sensitivity detection as a fluorimetric ion-pair reagent.
The use of LC-MS in which mass spectrometry is incorporated in HPLC as a detector has become widespread. Sodium alkanesulfonates, a general ion-pair reagent, being non-volatile crystals pose a problem in that they contaminate the interface. The IPC-PFFA series is made of highly volatile ion-pair reagents allowing for continuous LC-MS analysis without contaminating the interfaces.
저희는 사용자에 맞게 사이트를 설정하여 사용자가 사이트를보다 원활하게 이용할 수 있도록 쿠키를 사용합니다. 자세한내용은 개인 정보 보호 정책 을 참조하십시오. 사이트를 사용하면 쿠키 사용에 동의합니다
Notice of Discontinuing the Use of Password-Protected Compressed Files|Product Document Searching Made Easy by 2D Code! | TCI Chemistry News November 2024 | [Product Highlights] Hypervalent Iodine Reagent... | Various analytical charts can be searched on each product detail page and Product Document Search (The kinds of analytical charts differ by product)
Maximum quantity allowed is 999
Please select the quantity
Ion Pair Reagents (for HPLC)
Ion-exchange chromatography systems have previously been utilized in HPLC analysis of ionic samples. Reversed phase partition chromatography using ion-pair reagents has been developed and utilized. The ionic samples form an ion-pair with ion-pair reagents in the mobile phase to become electrically neutral. The increase in hydrophobic character of the ion-pair results in a greater affinity for the reverse stationary phase and leads to sample resolution.
Ion-Pair Chromatography for Acidic Samples
• Analysis is performed with pH adjusted to 7.5 with the addition of quaternary ammonium salts to the mobile phase.
• Acidic samples form an electrically neutral ion-pair with the quaternary ammonium salt and are retained in the reverse phase systems.
• The ion-pair reagents for acidic samples for LC-MS are supplied as 0.5M aqueous solutions and were adjusted to pH7.5. The solution can be used as a neutral mobile phase after dilution with the LC solvents (acetonitrile/water or methanol/water) to 5mM. Since the acidic substances are ionized under the neutral conditions, they are facilitated to form an ion-pair.
—Examples—
1) When 0.5mol/L Tetrabutylammonium Phosphate (IPC-TBA-P) [I0367] is used:
The reagent (10mL) is diluted to 1L with an aqueous solvent such as methanol - water. (pH adjustment is not required because I0367 is already buffered.)
2) When Tetrabutylammonium Hydroxide (IPC-TBA-OH) [I0364] is to be used:
i) The reagent (12.5mL) is diluted to 1L with an aqueous solvent such as methanol - water.
ii) The pH is adjusted to 7.5 by the addition of an aqueous phosphoric acid (10%).
• Analysis is performed with pH adjusted to 7.5 with the addition of quaternary ammonium salts to the mobile phase.
• Acidic samples form an electrically neutral ion-pair with the quaternary ammonium salt and are retained in the reverse phase systems.
• The ion-pair reagents for acidic samples for LC-MS are supplied as 0.5M aqueous solutions and were adjusted to pH7.5. The solution can be used as a neutral mobile phase after dilution with the LC solvents (acetonitrile/water or methanol/water) to 5mM. Since the acidic substances are ionized under the neutral conditions, they are facilitated to form an ion-pair.
—Examples—
1) When 0.5mol/L Tetrabutylammonium Phosphate (IPC-TBA-P) [I0367] is used:
The reagent (10mL) is diluted to 1L with an aqueous solvent such as methanol - water. (pH adjustment is not required because I0367 is already buffered.)
2) When Tetrabutylammonium Hydroxide (IPC-TBA-OH) [I0364] is to be used:
i) The reagent (12.5mL) is diluted to 1L with an aqueous solvent such as methanol - water.
ii) The pH is adjusted to 7.5 by the addition of an aqueous phosphoric acid (10%).
Ion-Pair Chromatography for Basic Samples
• Analysis is performed by the addition of sodium alkanesulfonate to the mobile phase.
• The basic samples form an electrically neutral ion-pair with sodium alkanesulfonate and are retained in the reverse phase system.
• In the case of sodium alkanesulfonate, the greater the number of carbons in the alkyl group, the greater the partition ratio.
• The solubility of the products such as sodium 1-decanesulfonate (IPC-ALKS-10) [I0348] may decrease depending upon the composition of the mobile phase solvents; especially after the addition of the buffer for pH adjustment. Resulted turbidity of the mobile phase and crystal formation may interfere with the analysis. To avoid the trouble, modification of the solvent system composition should be considered.
• The ion-pair reagents for basic samples in LC-MS analysis are supplied as 0.5M aqueous solutions. The solution can be used as an acidic mobile phase after dilution with the LC solvents (acetonitrile/water or methanol/water) to 5mM. Since the basic substances are ionized under the acidic conditions, they are facilitated to form an ion-pair.
• We supply the high-quality products (IPC-PFFA-6 HG [A5722], IPC-PFFA-7 HG [A5721], IPC-PFFA-8 HG [A5720]) of IPC-PFFA-6, IPC-PFFA-7 and IPC-PFFA-8 for high-sensitive detections.
—Examples—
1) Sodium 1-Heptanesulfonate (IPC-ALKS-7) [I0345] 1.011g (0.005mol) is weighed out.
2) The reagent is dissolved in 1L of an aqueous solvent such as methanol - water.
3) The pH is adjusted to 3.5 by the addition of aqueous phosphoric acid (50%).
• Analysis is performed by the addition of sodium alkanesulfonate to the mobile phase.
• The basic samples form an electrically neutral ion-pair with sodium alkanesulfonate and are retained in the reverse phase system.
• In the case of sodium alkanesulfonate, the greater the number of carbons in the alkyl group, the greater the partition ratio.
• The solubility of the products such as sodium 1-decanesulfonate (IPC-ALKS-10) [I0348] may decrease depending upon the composition of the mobile phase solvents; especially after the addition of the buffer for pH adjustment. Resulted turbidity of the mobile phase and crystal formation may interfere with the analysis. To avoid the trouble, modification of the solvent system composition should be considered.
• The ion-pair reagents for basic samples in LC-MS analysis are supplied as 0.5M aqueous solutions. The solution can be used as an acidic mobile phase after dilution with the LC solvents (acetonitrile/water or methanol/water) to 5mM. Since the basic substances are ionized under the acidic conditions, they are facilitated to form an ion-pair.
• We supply the high-quality products (IPC-PFFA-6 HG [A5722], IPC-PFFA-7 HG [A5721], IPC-PFFA-8 HG [A5720]) of IPC-PFFA-6, IPC-PFFA-7 and IPC-PFFA-8 for high-sensitive detections.
—Examples—
1) Sodium 1-Heptanesulfonate (IPC-ALKS-7) [I0345] 1.011g (0.005mol) is weighed out.
2) The reagent is dissolved in 1L of an aqueous solvent such as methanol - water.
3) The pH is adjusted to 3.5 by the addition of aqueous phosphoric acid (50%).
We also supply HPLC columns.